Calibration¶
With calibration, the scanner parameters are recalculated, which not only ensures the accuracy of the scanner, but also improve the quality of scanning.
Note
Calibration is required under the following conditions:
-
When the scanner is used for the first time.
-
The scanner was severely shaken or shocked, such as shocked during transportation.
-
Severe accuracy reduction, such as frequent errors in alignment or unrecognized markers.
-
Incomplete data is acquired during the scanning or serious deterioration of the quality of scanned data.
-
When the scanner is NOT calibrated more than 14 days, the software will prompt you to do calibration.
Warning
-
The calibration board is matched to the device. Doing the calibration with an incorrect calibration board will fail to generate good scan data or optimum accuracy.
-
Always make sure that both sides of the calibration board are clean and free of scratches.
-
Do not place heavy objects or sundries on the calibration board.
-
Keep the calibration board away from corrosives, metals and sharp objects to avoid corrosion or damage.
-
It is not recommended to wipe the calibration board. When cleaning the board becomes very necessary, gently wipe it with a piece of a clean damp cloth. Do not use a cloth with chemicals or alcohols to wipe the calibration board.
-
After using the calibration board, put it safely in a box or flannel bag.